Roche Accu-Chel Wartungshandbuch Seite 96

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 135
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 95
96
Measurement and storage conditions
Measurement and storage conditions11
So that the meter works reliably and you obtain accu-
rate results, make sure that the conditions listed on
the following pages are met.
11.1 Temperature
For blood glucose tests and performance checks
the temperature must be between +10 and
+40 °C.
If the temperature is at the limits of the permitted
range (between +5 and +10 °C or between +40
and +45 °C), you can still perform a test. The
thermometer symbol will, however, be displayed
(see also Chapter 12.1).
Do not use results obtained at the limits of
the permitted temperature range as a basis
for making therapeutic decisions, as they
could be incorrect. Incorrect results can
cause the wrong therapy recommendation
to be made and so lead to serious adverse
health effects.
Seitenansicht 95
1 2 ... 91 92 93 94 95 96 97 98 99 100 101 ... 134 135

Kommentare zu diesen Handbüchern

Keine Kommentare