A study was conducted and the precision of duplicates for capillary blood
results was calculated for both trained users and healthcare professionals. The
following results were obtained:
User Professional
Results Results
N 214 249
Mean 2.57 2.52
SD 0.13 0.13
CV 5.13 5.36
Precision:
109876543210
10
9
8
7
6
5
4
3
2
1
0
N = 297
PB: y = 1.154x - 0. 2
Slp CI (1.111,1.198)
Int CI (-0.3, -0.2)
Correlation = 0.934
109876543210
10
9
8
7
6
5
4
3
2
1
0
N = 463
PB: y = 1.000x + 0.0
Slp CI (1.000,1.030)
Int CI (-0.05, 0)
Correlation = 0.977
3 User Selection Guidelines User Selection Guidelines 3
This study shows that trained users are able to obtain results that are as precise as those
obtained by healthcare professionals trained in the use of the CoaguChek XS System.
During the study, the following error rates were observed:
Visit 1 Visit 2 Visit 3 Overall
Error 6 Test Strip Interference 0.8% 0.9% 0.5% 0.7%
Error 5 Blood Application 31.7% 23.9% 18.9% 25.2%
Error 4 Test Strip Unusable 0.8% 0.0% 2.8% 1.1%
Error 000 Time Exceeded 1.5% 1.3% 0.0% 1.0%
Error QC Quality Control Failure 1.1% 0.0% 0.0% 0.4%
Unknown Error 1.1% 0.0% 0.5% 0.6%