BLOOD GLUCOSE METER
User’s Manual
A
A
CCU-
CCU-
C
HE
HEK
®
Advantage
Seitenansicht 0
1 2 3 4 5 6 ... 89 90

Inhaltsverzeichnis

Seite 1 - Advantage

BLOOD GLUCOSE METERUser’s ManualAACCU-CCU-CHEHEK®Advantage

Seite 2

Special FeaturesThe meter is designed to store information important to you and your healthcare professional. The memory stores 480 results. Along wi

Seite 3

ixAbout the Accu-Chek Advantage PlusTest StripElectrodesInsert test strip with silver-coloured bars up and toward the meter.Yellow WindowDo not apply

Seite 5 - Contents

Inserting the BatteryThe meter is shipped with one 3-volt battery type 2032 that needs to be inserted before testing. The battery that comes with the

Seite 6

2. Insert the 3-volt battery (type 2032) into the compartment with “+” side facing you. 3. Snap the battery cover back in place. If the time and

Seite 7

If a flashing battery symbol appears on the display, the battery is nearly dead. After changing the battery, the meter prompts you to verify the time

Seite 8

Setting the Beeper, Time, Date, and Hypo FlagTurning the Beeper On or OffDoes the beeper have to be turned on?No, it is an optional feature. Turning t

Seite 9 - Code Chip Slot

3. Press and hold down the ON/OFF button until “set” appears at the bottom of the screen. Release the ON/OFF button. “On,” “set,” and the flashin

Seite 10 - Special Features

5. Press and hold down the ON/OFF button until the test strip symbol flashes or code and “---” appear to leave this function. Setting the Time and D

Seite 11 - Test Strip

5. Press and release the ON/OFF button. The hour display flashes.6. Press and release the right or left arrow button until the correct hour is dis

Seite 13 - Inserting the Battery

Setting the Hypo FlagThe hypo flag allows you to define your hypoglycemic level and receive a reminder whenever you obtain a blood glucose result belo

Seite 14 - Battery Symbol on the Display

5. Press and release the right or left arrow button to turn ON the hypo flag.6. Briefly press and release the ON/OFF button. “Hypo” flashes, and t

Seite 15 - Battery must be replaced

The code chip is properly inserted when it snaps into place. You need to insert a new code chip: • Whenever you open a new box of test strips. • W

Seite 16 - Hypo Flag

TestingGathering SuppliesTo begin testing, you need the following items: • Accu-Chek Advantage Meter• Accu-Chek Advantage Plus Test Strips• Accu-Chek

Seite 17

Any object coming into contact with human blood is a potential source of infection (see: Clinical and Laboratory Standards Institute: Protection of La

Seite 18 - Setting the Time and Date

5. Let your arm hang down at your side briefly to allow blood to flow to your fingertips.6. Grasp your finger just below the joint closest to th

Seite 19

9. When blood is applied to the test strip, an hourglass symbol flashes on the display until the measurement is completed. The result appears on the

Seite 20 - Setting the Hypo Flag

11. Remove the test strip from the meter. Discard the used test strip and lancet in a puncture-proof container with a lid. If you have completed your

Seite 21 - Coding the Meter

Marking a Result with a Flag, Asterisk ( )This feature allows you to perform a test and mark the result with a flag (asterisk) to be stored in memory.

Seite 22

Normal Blood Glucose ValuesThese test strips are calibrated to deliver whole blood results. The normal fasting blood glucose range for an adult withou

Seite 23 - Performing a Test

WelcomeThank you for selecting the Accu-Chek Advantage System. Designed for ease of use, this system was produced in close association with diabetes h

Seite 24

Then perform a quality control check with glucose control solutions and a new test strip. (See “Quality Control Checks.”) If the control result is wit

Seite 25

Correct Test Strip AppearanceAlthough only a small amount of blood is needed, it is very important that you put enough blood on the test strip so that

Seite 26

Using the MemoryThe meter automatically stores 480 glucose results, letting you review them in order from the most recent to the oldest. If you have s

Seite 27

chip is installed, the code number appears briefly, followed by the most recent result stored in memory. If no code chip is installed, code and “---”

Seite 28

Quality Control ChecksWhat are quality control checks?They ensure that you are doing a test correctly and that your system is working properly. When s

Seite 29 - Normal Blood Glucose Values

To do a quality control check, you need: • Accu-Chek Advantage II Glucose Control Solutions (Level 1 and Level 2) • Accu-Chek Advantage Plus Test S

Seite 30 - Do’s and Don’ts

Performing a Quality Control Check 1. Open the test strip container and remove a test strip. Replace the container cap. 2. Within three minutes,

Seite 31

5. Gently squeeze the bottle to form one small drop.6. Bring the drop to the edge of the test strip within the curve, and allow the test strip to a

Seite 32 - Using the Memory

Note: Result must be marked with an L1 or L2 for the meter to verify if the control result is within range. If result has not been marked with an L1 o

Seite 33

Understanding Control ResultsIf your glucose control results are within the acceptable range, you can feel confident that the test strips and meter ar

Seite 34 - Quality Control Checks

The Accu-Chek Softclix finger pricker is intended for patient self-monitoring by a single person only. It must not be used to collect blood from more

Seite 35

Note: Consumers – please proceed to the “Caring for Your Meter” section. The following sections are for healthcare professionals only.Patient Testing

Seite 36

Testing Blood Glucose in Multiple PatientsOnly healthcare professionals are allowed to perform blood glucose tests in a multiple patient testing situa

Seite 37

The Accu-Chek Softclix finger pricker is intended for patient self-monitoring by a single person only. It must not be used to collect blood from mor

Seite 38

4. Use a finger pricker with adjustable penetration depth settings. If using a Safe-T-Pro Plus finger pricker, select a penetration depth to suit the

Seite 39 - Understanding Control Results

32Flagging Test Results It is possible to flag a test result with an asterisk ( ) for a special result. When the results in memory are reviewed, this

Seite 40 - Sample handling

33 Dispose of used lancets, test strips or disposable finger prickers according to local health and safety regulations.Note: The test strip cannot pr

Seite 41

Troubleshooting Checks1. Check if the test strips were expired.2. Check if the cap on the test strip container was always closed tightly.3. Check if

Seite 42

4. Repeat the test with a properly stored test strip.5. Read “Patient Testing for Healthcare Providers” and test again. If you still have problems, c

Seite 43

36How to Disinfect• Make sure the meter is off.• Gently wipe the meter’s surface.• Cotton buds, wipes and cloths lightly moistened with disinfectan

Seite 44 - Flagging Test Results

37If blood has entered the meter, wipe the area around the test strip slot and the display screen. Allow areas you have wiped time to dry thoroughly.T

Seite 45 - Unusual Test Results

ContentsAbout the Accu-Chek Advantage Meter . . . viSpecial Features . . . . . . . . . . . . . . . . . . . . viiiAbout the Accu-Chek Advantage Plus T

Seite 46 - Troubleshooting Checks

38Control Testing for Healthcare ProfessionalsIntroductionTesting control solutions with known glucose levels establishes that the operator and the sy

Seite 47 - What to Disinfect

39Control results must be within the designated range on the test strip container label, or as defined by your facility, before being considered accep

Seite 48 - When to Disinfect

403. Press and release the right arrow button once to select a L1 (Level 1) control or twice to select a L2 (Level 2) control.Note: Results marked as

Seite 49

41Example of an acceptable Level 1 control result.If result has not been marked with an L1 or L2, refer to the test strip container label for the gluc

Seite 50 - Testing Intervals

42Caring For Your MeterDisplay CheckThis verifies that all parts of the display are working. Make sure the meter is off. Press and hold down the ON/OF

Seite 51 - Performing a Control Test

43Storage and HandlingKeep the meter free of dust. Protect it from extremes in temperature and humidity, and handle it with care. If the meter is drop

Seite 52

Quick Reference Display GuideMeter needs to be coded or code chip is not inserted completely; turn meter OFF and insert code chip. Meter is in the “se

Seite 53

Blood glucose result appears on display. Blood glucose result may be higher than 33,3 mmol/L. If this does not confirm the way you feel, review proper

Seite 54 - Caring For Your Meter

Temperature is above or below the operating range of test strips (14 ° to 40 °C). The result you have obtained may not be accurate. Move to an area be

Seite 55 - Transferring Data

If you are reviewing memory and a result appears with a temperature symbol as shown, this result may not be accurate because the test was performed ou

Seite 56 - Quick Reference Display Guide

Patient Testing for Healthcare Professionals . . . . . . . . . . . . . . . . . . . . . . . 28 Sample Handling . . . . . . . . . . . . . . . . . . .

Seite 57

Display MessagesBattery DisplaysBattery power is low. Change battery soon. (battery symbol not flashing)Battery is dead. Change battery now. Meter tu

Seite 58 - Memory Displays

Quality Control Check DisplaysMeter is ready for a Level 1 or Level 2 glucose control check. Apply a drop of Level 1 or Level 2 glucose control soluti

Seite 59

Example of an acceptable control result.Glucose control result is not acceptable. Repeat test. Refer to Understanding Control Results (see page 27).Ex

Seite 60 - Display Messages

Error MessagesThe meter automatically performs a system check to prevent inaccurate results. If the meter detects a malfunction, it displays the “Err”

Seite 61

Your blood glucose may be extremely high or a meter or test strip error has occurred. If this confirms how you feel, contact your healthcare professio

Seite 62

Meter has detected the loss of all results in memory. Turn meter OFF and turn it back ON again. You may not be able to recall the results previously s

Seite 63 - Error Messages

Technical DataPlease read information packaged with test strips regarding up to date product specifications and limitations.Measuring Principle: The g

Seite 64

Product SpecificationsBlood Volume 4LSample Type Fresh capillary whole bloodMeasuring Range System measuring range is 0.6–33.3 mmol/L.Test Strip

Seite 65

Construction Hand heldProtection Class IIIRated Operation Continuous operationData Port 3-wire RS-232C Serial Port Electromagnetic CompatibilityT

Seite 66 - Technical Data

The Accu-Chek Advantage System complies with the requirements of EN ISO 15197.References1. Stedman, TL. Stedman’s Medical Dictionary, 27th Edition, 1

Seite 67 - C and less than 85% humidity

Error Messages . . . . . . . . . . . . . . . . . . . . . 51Technical Data. . . . . . . . . . . . . . . . . . . . . . 54Product Specifications . .

Seite 69 - Safety Remarks

About the Accu-Chek Softclix Finger Pricker A Cap B Depth selection arrow C Notch D Protective cap E Accu-Chek Softclix lancet F Lancet holder G

Seite 70

Accu-Chek Softclix Finger PrickerIntroductionThe Accu-Chek Softclix finger pricker is an easy-to-use finger pricking device for obtaining blood from t

Seite 71

For this purpose the Accu-Chek Softclix finger pricker can be set to different penetration depths ranging from 0.5 to 5.5. They are indicated on the c

Seite 72

23 4576162

Seite 73

How to Take a Blood Sample1. Pull off the cap from the Accu-Chek Softclix finger pricker.2. Insert a new Accu-Chek Softclix lancet as far as it goes i

Seite 74

Wait a few seconds to allow the puncture site to open so that a drop of blood can form. Apply the blood drop to the test strip. Then carry out the t

Seite 75 - How to Take a Blood Sample

Subsequent Treatment of the Puncture SiteWipe the puncture site with a dry, clean tissue. If it continues to bleed, apply gentle pressure to the punct

Seite 76 - Selecting the Puncture Site

Product Safety Information Strong electromagnetic fields may interfere with the proper operation of the meter. Do not use this meter close to source

Seite 77 - Accu-Chek Softclix Lancet

Please consult instructions for use Caution (refer to accompanying documents). Please refer to safety-related notes in the manual accompanying

Seite 78 - Explanation of Symbols

viAbout the Accu-Chek Advantage Meter DisplayShows blood glucose result, messages, and glucose results stored in memory.ON/OFF ButtonPress to turnmet

Seite 79

Dispose of in domestic waste 3-volt lithium battery (coin cell 2032)Accu-Chek Softclix Finger PrickerOn the finger pricker itself and on the pa

Seite 80

Warranty and Service InstructionsRoche Diagnostics warrants the meter against any material and manufacturing defects that may arise during proper use

Seite 81 - Supplies and Accessories

Accu-Chek Softclix Finger Pricker GuaranteeThe Accu-Chek Softclix Finger Pricker is guaranteed for life from the date of purchase against defects in m

Seite 82 - Guarantee

IndexAbout the Accu-Chek Softclix Finger Pricker ... 59About the Accu-Chek Advantage Meter ... viAccu-Chek Advantage Plus

Seite 83

72Error messages ...51Flag (asterisk) ...

Seite 84

73Testing, blood glucose ... 11Testing, quality control ... 22, 49Te

Seite 85

74Any QuestionsWe want to give you the help you need for long and reliable service of your meter. Review your test procedure and make sure that you ar

Seite 88

This product fulfills the requirements of Directive 98/79/EC on in vitro diagnostic medical devices.

Seite 89

viiAbout the Accu-Chek Advantage MeterCode ChipFor coding the meter. A new code chip comes with every container of test strips.Battery CoverRemove cov

Seite 90 - 68298 Mannheim

ACCU-CHEK, ADVANTAGE, and SOFTCLIX are trademarks of Roche.Accu-Chek Customer Careline*0800 701 000 (UK)1 800 709 600 (Ireland)*Calls may be recorded