Proficiency Checks
5
After completing the training class and beginning self-testing, users should be
monitored periodically by a trained healthcare professional to be sure their testing
technique is correct and the meter is performing properly. Suggested frequency
for monitoring is every six months. The following procedure may be useful:
Establish a reminder system that prompts you to contact each user at
6-month intervals.
Make an appointment for the user to come to the clinic or office.
Ask the user to bring the CoaguChek XS meter, testing supplies, and
Log Book.
During the interview, record your observations on the 6-month
Performance Evaluation form.
Inspect the meter thoroughly.
Observe the user cleaning the CoaguChek XS meter. If necessary, stress
the importance of maintaining a regular cleaning schedule.
Observe the user performing a blood test. If necessary, offer
suggestions for improvement. A 2-week follow-up is recommended
whenever competency needs reassessment.
The healthcare professional will perform a user blood test after the
trainee performs a self-test. The test should match within 30%; this
verifies that the meter is performing properly. If this test indicates a
problem with the meter or test strips, repeat the test. If the result is still
in question, please contact the Roche Diagnostics Technical Service
Center 1-800-428-4674, 24 hours a day, 7 days a week, 365 days a year.
If the test result is within range and the user performs all tests
satisfactorily, make an appointment for six months in the future.