Roche Accu-Chek Aviva Technical Information Seite 228

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 282
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 227
226
7.5 Maintenance and Testing
The meter needs little or no maintenance with normal use. It automatically tests its own systems every
time you turn it on and lets you know if something is wrong. For detailed information on troubleshooting the
meter, see Chapter 8, “Troubleshooting.”
If you drop the meter or think it is not giving accurate results, ensure your test strips and control solution
have not expired, and then perform a control test.
• Perform a control test with each new vial of test strips.
To test the meter display, turn off the meter, and then press and hold
+. The display cycles through colours
(red, blue, green, and white). If any part of the display does not change colours, contact Roche.
• If the meter is not working properly, contact Roche.
50813_00048009694_D.indb 226 7/3/12 3:13 PM
Seitenansicht 227
1 2 ... 223 224 225 226 227 228 229 230 231 232 233 ... 281 282

Kommentare zu diesen Handbüchern

Keine Kommentare