Display Shows Possible Cause(s) Possible Solution(s)
Above Hyper
Warning Limit
j Your test result is above the
hyper warning limit set in the
meter.
j Treat your high blood glucose as recommended by
your healthcare professional.
j Consider checking bG, ketones, and insulin.
j See the pump User Guide for additional solutions.
Beep/Vibrate O j Beep and Vibrate settings
cannot be turned o at the
same time.
j Ensure that either the Beep or Vibrate setting is
turned on. See Chapter 5, “Changing Meter
Settings.”
Below Hypo
Warning Limit
j Your test result is below the
hypo warning limit set in the
meter.
j Treat your low blood glucose as recommended by
your healthcare professional.
j The meter displays a recommended number of
carbohydrates for you to eat, and then retest your
blood glucose.
Bolus Advice
Bolus Advice data
deleted
j Bolus advice data integrity
cannot be confirmed.
j Wait 8 hours for accurate bolus advice.
IMPORTANT:
• Insulin doses and meals taken before the Bolus Advice warning are no longer
reflected in the bolus advice calculation.
• The meter may not reflect the pump bolus history, however the bolus history is
available on the pump.