Cleaning/Disinfecting the Test Strip Guide
n Use only 70% isopropyl alcohol or 10% bleach solution to clean/disinfect the CoaguChek
Plus test strip guide.
n Do not use any other cleaning/disinfecting solutions on the test strip guide.
Use of other cleaning/disinfecting solutions could result in damage to the meter.
Cleaning
84
1 Open the cover
With the meter powered off, use your
thumbnail to open the cover of the test
strip guide by pressing its front edge
upward. Move the cover safely away from
the meter. Then rinse the cover with water
or wipe it clean.
2 Clean the test strip guide
n Hold the meter upright with the test
strip guide facing down.
n Clean the easily accessible areas with a
cotton swab.
n Ensure the swab is only damp, not wet.
n Apply cleaning agent for a contact
time of >1 minute (refer to the
corresponding product labeling).
n Wipe away residual moisture and fluids.
Caution: Do not insert any objects
into the test strip guide. Doing so could
damage the electrical contacts behind the
test strip guide.
Please note: Let the inside of the test
strip guide dry for at least 10 minutes
before re-attaching the test strip guide
cover and testing again.