Error messages
93
Error Description/Solution
Error: Quality Control
This message appears if quality control lockout was set up for
NEW CODE lockout.
A quality control test must be carried out when a new test
strip lot is used. The quality control for new lot is missing.
Error
QC
Solution
Perform a quality control test with the new test strip lot.
Error: Measurement
Measurement error resulting from the blood sample
Error
7
Solution
Turn the meter off and remove the test strip. Repeat the
measurement using a new test strip and blood taken from a
new fingerstick. Do not touch or remove the test strip when a
test is in progress. Be sure to apply the blood drop within 15
seconds of lancing the fingertip.
Note on Error 7 “Measurement”: The CoaguChek XS PT test strip may be used for
patients under a combination therapy of oral anticoagulants plus heparin injections. For
maximum heparin concentrations that do not interfere with the test, please refer to the
package insert. Under no circumstances, however, should heparinized capillary tubes be
used for sample application. If capillary tubes are used, please use only the dedicated
CoaguChek capillary tubes. Be sure to apply the blood drop to the test strip within 15
seconds of lancing the fingertip.
In rare cases, patients with long clotting times (>8 INR) may receive an
“ERROR 7” message on the meter display. If this error message appears again
when the test is repeated, the result must be checked using another method.
Under no circumstances should heparinized capillary tubes be used. Please
use only the dedicated CoaguChek capillary tubes.