CoaguChek® XS PlusSystemUser Manual

Seite 2

Operating conditionsTo be sure that your CoaguChek XS Plus System functions properly, please observe the following guidelines:n Only use the meter at

Seite 3

Further information100this section to claim rights or contest your rights to work written entirely by you; rather, the intent is to exercise the right

Seite 4 - The CoaguChek XS Plus System

Further information1015. You are not required to accept this License, since you have not signed it. However, nothing else grants you permission to mod

Seite 5

Further information1029. The Free Software Foundation may publish revised and/or new versions of the General Public License from time to time. Such ne

Seite 6

Further information103How to Apply These Terms to Your New Programs If you develop a new program, and you want it to be of the greatest possible use t

Seite 7 - Introduction

Further information104This software was developed by the Computer Systems Engineering group at Lawrence Berkeley Laboratory under DARPA contract BG 91

Seite 8 - Contents of the pack

Operating conditions and technical dataTemperature range 59°F to 90°F (15°C to 32°C)Relative humidity 10% to 85% (no condensation)Maximum altitude

Seite 9 - Protection against infection

Sample materialSample type Capillary whole blood or non-anticoagulated venous whole blood.Sample size At least 8 µLInteractions Refer to test stri

Seite 10 - Electromagnetic interference

CoaguChek XS Plus Meter Limited (1 Year) WarrantyROCHE DIAGNOSTICS warrants, to the original purchaser only, that the meter shall be free from all def

Seite 11 - Quality control

0123 COAGUCHEK is a trademark of Roche. Manufactured for: Roche Diagnostics 9115 Hague Road Indianapolis, IN 46256 Roche Diagnostics GmbH Sand

Seite 12 - About this manual

Quality controlThe CoaguChek XS Plus System has a number of built-in quality-control functions.n A check of the electronic components and functions e

Seite 13

Manual formatThe layout of this operator’s manual has been specially designed to help you locate the most important information quickly and easily. Al

Seite 14 - The CoaguChek XS Plus meter

Example 2: This column contains an illustration of the This column contains information linked to display screen. the display screen. The main m

Seite 15

The CoaguChek XS Plus meterThe CoaguChek XS Plus meter14ABCEFJFHIGD

Seite 16 - Power supply

Meter elements overview A Touch screen Shows results, information, symbols and results recalled from memory. To select any of the screen prompts,

Seite 17 - Operating the instrument

Power supplyThe CoaguChek XS Plus meter can operate with either the power adapter provided, batteries, or a special rechargeable battery pack (REF 048

Seite 18 - Inserting the batteries

Before using the meter for the first time, carry out the following steps:1 Insert batteries* and/or connect the power adapter.2 Set the current date a

Seite 19 - Meter setup

Inserting the batteriesOperating the instrument181 With the meter turned off, press the battery compartment cover release tab and slide the cover o

Seite 20 - Symbol Meaning

Meter setup You can call up any displayed function by touching the screen prompt for it with your finger (or a special pen for this purpose). Tou

Seite 21

The contents of this manual, including all graphics and photographs, are the property of Roche Diagnostics. No part of this document may be reproduced

Seite 22

Meter setup20Display field and symbolsThe screen prompts that appear during normal operation are shown here, along with their meaning. Error messages

Seite 23 - Settings summary

Meter setup21Symbol Meaning180 SECApply blood sample (the time left to apply sample is counted down in the display)180 SECApply QC sample (the time le

Seite 24

22Meter setupSymbol MeaningBattery status: • When the batteries still have their full charge, all segments are lit.• Individual segments disappear

Seite 25

Meter setup23Settings summaryThe diagram below shows all of the SETUP areas that can be accessed on the meter.SetupScreenLanguageUnitsContrast Date/Ti

Seite 26

Meter setup24 Group Subgroup Setting Values * SCREEN CONTRAST 0 - 10 (5 *) RESULT UNITS INR * INR/SEC INR/%Q LANGUAGE Dansk

Seite 27

Meter setup25 Group Subgroup Setting Values * OPTIONS SORT DATE/TIME * PATIENT ID NUMBER BEEPER BEEPER OFF LOW MEDIUM *

Seite 28 - Contrast

Meter setup26 Group Subgroup Setting Values * ID SETUP Admin. (Administrator) Blank (off) * OPERATOR (Operator list is optional) Blank (off) *

Seite 29

Meter setup27Conventions used in this manualThis manual refers to all the various operations of the meter. When discussing a particular operation, a p

Seite 30 - Language Selection

Screen setupThe Screen setup area contains the options for changing the way the display looks. ContrastUse this setup option to make the screen displa

Seite 31

Meter setup29UnitsUse this setting to select the unit(s) of measure to be displayed with the results.1 From the main menu screen, touch Setup.2 Fr

Seite 32 - Setting the date

Revisions to this document are provided by Roche Diagnostics when necessary. No part of this document may be reproduced in any form or by any means

Seite 33 - Setting the time

Meter setup30Language SelectionUse this setting to select the language for all displays (that contain text).1 Touch Setup in the main menu.2 From

Seite 34

Meter setup314 Touch or to scroll to the language of choice on the screen. If the arrow is just an outline , this means the arrow is not active

Seite 35 - Options setup

Meter setup32Setting the dateWhen you turn on the meter for the first time (or after a long period without power), the input field for the date automa

Seite 36

Meter setup33Setting the timeWhen you turn on the meter for the first time (or after a long period without power), the Setup menu appears automaticall

Seite 37 - Key Click:

Meter setup34Setting the display options for date and timeYou don’t have to select display options for date and time, even though you must enter the

Seite 38 - Auto off

Meter setup35Options setupSortSort refers to the order in which measured and stored results are displayed when you use your CoaguChek XS Plus meter’s

Seite 39 - Computer

BeeperNote: Your CoaguChek XS Plus meter can display information visually and alert you to special circumstances with a beeper. When the beeper is act

Seite 40 - ID Setup

Meter setup374 Touch the screen prompt with the desired setting for the beeper, then touch the prompt with the setting of choice for the key click.

Seite 41 - System Administrator (Admin.)

Meter setup38Auto offYou can set up your CoaguChek XS Plus meter so that it turns itself off automatically if it has not been used (no buttons pressed

Seite 42

Meter setup39ComputerYour CoaguChek XS Plus meter can connect with a computer via an infrared interface using optional base unit. Using a computer, yo

Seite 43 - Changing an existing Admin

The CoaguChek XS Plus SystemThe CoaguChek XS Plus System is used to monitor coagulation (blood-clotting) values (prothrombin time, PT, Quick value) wi

Seite 44 - Admin. password:

ID SetupUse the ID Setup screen to enter settings for user management and patient management. These settings are optional; the meter can be operated w

Seite 45 - If Operator ID numbers are

Meter setup41System Administrator (Admin.)In the default setting, the meter is not protected with an Admin. password, and all setup options are access

Seite 46

Meter setup424 Using the keypad displayed on the screen, enter the Admin. password of choice. The password should include up to 20 characters and/or

Seite 47 - Alphanumeric

Meter setup431 From the main menu screen, touch Setup.2 From the Setup menu screen, touch ID Setup.3 Using the keypad displayed on the screen,

Seite 48 - Lockouts setup

Meter setup441 From the main menu screen, touch Setup.2 From the Setup menu screen, touch ID Setup.3 Using the keypad displayed on the screen,

Seite 49

Operator (only available with optional software)To specify operators, an operator list must be created on the PC and downloaded into the meter using s

Seite 50 - Number of Levels

PatientIn the default setting, input of Patient ID is “optional.” This means a Patient ID can be entered for each test, but this isn’t required. If yo

Seite 51 - QC (Quality control) lockout

Meter setup476 Select the input format of the Patient ID to be entered.7 Touch the screen prompt with the format of choice for the Patient ID. You

Seite 52 - Tw o Levels

Lockouts setupNote: The Lockouts menu contains the options for liquid quality control that require the operator to perform a liquid quality control te

Seite 53 - Important notes

Operator lockout (only available with optional software)To perform a test with the CoaguChek XS Plus meter, the individual steps must be performed pro

Seite 54

Introduction ...7 The CoaguChe

Seite 55 - Preparing to test

Meter setup505 Touch or to display the option of choice on the screen. If the arrow is just an outline , this means the button is not active.6

Seite 56 - Inserting the code chip

QC (Quality control) lockoutQC Lockout ensures that the optional liquid quality control tests are run on a regular basis for the meter to be used. Thi

Seite 57

Meter setup525 Touch the screen prompt with your option of choice for changing the test strip lot.6 Touch the screen prompt to select the interval

Seite 58 - Switching on the meter

Gather what you need:n CoaguChek XS Plus metern CoaguChek XS PT Test Strips and matching code chipn CoaguChek Lancetn CoaguChek Capillary tube/bulb

Seite 59

Getting a good capillary blood sampleTo get a suitable drop of blood:n Warm the hand. Have the patient hold it under his or her arm, use a hand warme

Seite 60 - Performing a test

Preparing to testTesting a capillary blood sample551 Have the test strip container at hand.2 Make sure that the code chip supplied with these test

Seite 61

Test strip code chipThe test strip code chip provides the meter with important information that it needs to perform the coagulation test. The chip con

Seite 62

Always compare the code number you see on the display with the number that is printed on the test strip container you are using. If the two code numbe

Seite 63

Switching on the meterWithout operator listTesting a capillary blood sample581 Place the meter on a level, vibration-free surface, or hold it in you

Seite 64

With operator list (only available with optional software)Testing a capillary blood sample593 If an operator list exists, wait until this list is di

Seite 65

Testing a capillary blood sample ... 53 Important notes ...

Seite 66

Testing a capillary blood sample60Performing a test8 Check the battery level. If there are no bars left in the battery symbol, you cannot perform an

Seite 67

Without patient listTesting a capillary blood sample6110 Touch Patient Test.11 If the Patient ID option was chosen as either optional or required wh

Seite 68

With patient list (only available with optional software)Testing the capillary blood sample6212 Touch Patient Test.13 Select the patient to be teste

Seite 69

Testing a capillary blood sample6315 The test strip symbol prompts you to insert a test strip. Remove a test strip from its container and close the c

Seite 70

Testing a capillary blood sample64Test09:15 am09/16/2009P-ID: PID01Code: 184Smith, HenryOp: Operator 1Par: PTThe hourglass symbol shows that the test

Seite 71

Testing a capillary blood sample6518 Now lance the side of a fingertip with the lancing device.19 Apply the blood directly from the finger to the se

Seite 72

Testing the capillary blood sample66Test09:15 am09/16/2009P-ID: PID01Code: 184Smith, HenryOp: Operator 1Par: PTDo not add more blood. Do not touch the

Seite 73

Testing a capillary blood sample67Test09:15 am09/16/2009P-ID: PID01Code: 184Smith, HenryOp: Operator 1Par: PTQCTest09:15 am09/16/2009P-ID: PID01Code:1

Seite 74 - Code: 184

Testing a capillary blood sample68Remove09:15 am09/16/200920 After the test and results are displayed, a Remove menu will appear on the screen. Remov

Seite 75

The CoaguChek XS Plus System has a number of built-in quality-control functions.n A check of the electronic components and functions every time the m

Seite 76

The CoaguChek XS Plus SystemThe CoaguChek XS Plus System measures blood-clotting time (prothrombin time) for people who are taking anticoagulation me

Seite 77

Preparing to run a liquid quality control testPrepare for a liquid quality control test in the same way you would prepare to perform a test with a ca

Seite 78

Quality control717 Close the container again.8 Make sure the dropper is at hand for the next steps in the liquid quality control test.9 Swirl th

Seite 79 - Review Results

Performing a liquid quality control testQuality control7210 Place the meter on a level, vibration-free surface, or hold it in your hand so it is roug

Seite 80 - Patient Result

Quality control7315 Touch Control Test.16 The test strip symbol prompts you to insert a test strip. Remove a test strip from its container and close

Seite 81 - Display Patient Result memory

Quality control74QC Test09:15 am09/16/2009184Op: Operator 1QC Test09:15 am09/16/2009QC 248QC 379QC 462NewQC 332*QC Test09:15 am09/16/2009Code: 184Op:

Seite 82

Quality control75QC Test09:15 am09/16/2009QC: 999Code: 184Op: Operator 1Par: PTLevel 1Level 2QC Test09:15 am09/16/2009QC: 999Code: 184Op: Operator 1Pa

Seite 83 - Cleaning/Disinfecting

Quality control7621 Using the dropper, draw up the dissolved contents of the vial. 22 Apply a single drop of control solution directly from the drop

Seite 84

Quality control77QC Test09:15 am09/16/2009QC: 999Code: 184Op: Operator 1Par: PTLevel: 1QC Test09:15 am09/16/20091.2 INR09/16/2009 09:15 amQC: 999Code:

Seite 85

If a liquid quality control test fails, an up arrow (too high) or down arrow (too low) is displayed and flashes.In this display, you can scroll throug

Seite 86 - Error messages

The CoaguChek XS Plus meter has a 1,000-value memory for saving results (patient tests and/or liquid quality control tests) together with time and dat

Seite 87

Test principleThe CoaguChek XS PT Test, used as directed with the CoaguChek XS Plus Meter, will provide an electrochemical measurement of prothrombin

Seite 88

Review Results80Memory09:15 am09/16/2009QC ResultPatient Result4 Select the function you want to perform in the memory.The following symbols for gen

Seite 89

Display Patient Result memoryReview Results811 Touch or to display the entry of choice on the screen. If the arrow is just an outline , this mea

Seite 90

Display QC (Quality Control) memoryReview Results82Memory09:15 am09/16/2009PID0109/14/2009 10:10 amPID0109/13/2009 10:07 amPID0109/12/2009 10:01 amPID

Seite 91

It is important to follow the procedures below to clean and disinfect the meter. Failure to follow these procedures may cause malfunction of the meter

Seite 92

Cleaning/Disinfecting the Test Strip Guiden Use only 70% isopropyl alcohol or 10% bleach solution to clean/disinfect the CoaguChek Plus test strip gu

Seite 93

Cleaning853 Allow to dry With the cover off, allow the test strip guide to dry for about 10 minutes.4 Close the cover Close the cover, and make sure

Seite 94

Depending on the circumstances, an error message may appear on the display of your meter.Error messages can have up to three warning levels as indicat

Seite 95

Error messages87Error message (always prefaced by Error)Brief descriptionErrorTest strip guide cover openErrorJVKLCode chip errorErrorEXPTest strip be

Seite 96 - DD:MM:YY

Error messages88Error message (always prefaced by Error)Brief descriptionErrorQCQuality control: Test strip unusable or quality control for new lot mi

Seite 97 - Further information

Error messages89Error message (always prefaced by Error)Brief descriptionErrorData transfer errorErrorEXPControl solution beyond expiration dateErrorM

Seite 98

Safety InformationThe CoaguChek XS Plus System9Protection against infectionMedical staff using the CoaguChek XS Plus System to perform tests on patien

Seite 99

Error messages90Error Description/SolutionError: Test stripA test strip was already inserted when the meter was turned on with the ON/OFF button.Error

Seite 100

Error messages91Error Description/SolutionError: CoverThe test strip guide cover is not properly closed.ErrorSolutionClose the test strip guide cover.

Seite 101

Error messages92Error Description/SolutionError: Time for application exceededThe maximum allowed time of 180 seconds for applying blood has been exce

Seite 102

Error messages93Error Description/SolutionError: Quality ControlThis message appears if quality control lockout was set up for NEW CODE lockout.A qual

Seite 103

Error messages94Error Description/SolutionError: Diagnostic testError during the internal diagnostic test which the meter performs.Error8SolutionTurn

Seite 104

Error messages95Error Description/SolutionError: Memory test (setup)The memory contains faulty data.ErrorSolutionThe meter has a fault. Please call Ro

Seite 105 - Product specifications

96Error messagesError Description/SolutionError: Date/time settingDate/time is inconsistentErrorDD:MM:YY?SolutionCheck date/time.Error: PasswordThe en

Seite 106 - Support/Services

97OrderingPlease contact Roche Diagnostics Customer Service at 1-800-428-5076.CoaguChek XS PT Test, 2 x 24 test strips REF 04625315160CoaguChek XS Pl

Seite 107 - Warranty

PreambleThe licenses for most software are designed to take away your freedom to share and change it. By contrast, the GNU General Public License is i

Seite 108 - 0 5021464001 (02) 2009-11 USA