Power supply
The CoaguChek XS Plus meter can operate with either the power adapter provided, batteries,
or a special rechargeable battery pack (REF 04805640001). The CoaguChek XS Plus System
uses four AA batteries. The batteries should last for about 60 tests, depending on the type of
battery used. You should insert the batteries or optional rechargeable battery pack even when
you use the power adapter. This ensures that you will not lose the date and time settings if
the power goes out.
To save power, the CoaguChek XS Plus meter has the option to automatically turn itself off
based on your setup selections, unless a screen prompt has been pressed or a new test strip
has been inserted. The default setting is set to OFF. When the meter turns itself off, all results
obtained up to that point remain in memory and the settings will still be there when you
power the meter back on. (Refer to Auto off in the chapter entitled Options setup.)
During battery operation, the meter always
displays the battery power level.
The meter retains results in memory even when no batteries are inserted. All other settings,
other than date and time after 10 minutes, are retained as well.
The CoaguChek XS Plus meter
16
Main Menu
09:15 am
09/16/2009
Control Test
Review Results
Setup
Patient Test
When replacing the batteries, you must
insert the new batteries within 10 minutes
of removing the old ones, to keep the
date and time settings. If you take longer
than this, you may need to re-enter the
date and time. To make sure that you
do not lose your date and time settings,
connect the power adapter while you
change the batteries.