Initial Startup
114
Options for glucose control tests
This menu contains options that allow you to specify
whether and at what intervals glucose control tests are
performed. If you require control tests, check your
set tings for STAT tests (see page 111).
1 Touch the desired option to enable it:
n Control Frequency
Always OK: QC Lockout is disabled, patient tests
can be performed at any time. This applies
regardless of the result of running glucose control
tests.
Last QC OK: QC Lockout is activated only when the
last glucose control test is out of range or a control
test is not available (e.g., when changing the test
strip lot).
Hours: QC Lockout is activated after the specied
time. When a glucose control test is run
successfully, the counter is reset to “0”.
Strip Count: QC Lockout is activated after reaching
the specied number of tests. When a glucose
control test is run successfully, the test strip
counter is reset to “0”.
Shift: QC Lockout is activated per shift.
Time Of Day: QC Lockout is activated at the
specied time of day. (Up to six times per day can
be congured.)
Rotating QC: QC tests require alternating use of L1
and L2 controls, instead of requiring both controls
when performing a QC test.
2 Touch
to save the settings. If the Hours or Strip
Count option is enabled, enter the desired number,
or:
3 Touch
to exit this menu without saving any
changes.
The display returns in both cases to the Setup Menu.