Introduction
19
Touchscreen
NOTICE
n
Use only your nger to touch the screen elements.
Using a sharp-edged object (e.g., tip of a pen) can
damage the touchscreen.
n Do not use the system in direct sunlight. Direct
sunlight may reduce the life expectancy and
functionality of the display, as well as the integrity of
test strips.
Laser scanner
The built-in barcode scanner emits a laser beam when
activated.
The built-in barcode scanner is a Class 1 laser, according
to IEC 60825-1/A2:2001.
WARNING
A barcode does not need to be present for the laser
scanner to become active. Do not stare directly into the
laser beam.
Electrostatic discharge (ESD)
The ACCU-CHEK Inform II system complies with both the
electromagnetic immunity requirements and radio
inter ference immunity requirements at the frequency and
test levels according to EN ISO 15197 Annex A. The
system also complies with the electrostatic discharge
(ESD) immunity requirements as specied in IEC
61000-4-2 and complies with applicable electromagnetic
emissions (EMC) requirements according to EN 61236.
The system’s emissions of the energy used are low and
not likely to cause interference in nearby electronically
driven equip ment.
Electrostatic discharge (ESD) is an electrical charge at
rest, most commonly known as static electricity. If the
ACCU-CHEK Inform II meter experiences a form of ESD
before or after running a blood glucose test, the test
result will be stored in the meter memory and transmitted
when the meter is docked in a connected base unit or
when a wireless connection is established. If the meter
experiences ESD during a blood glucose test, an error
message is displayed and no blood glucose result is
dis played by the meter or stored in the meter’s memory.
The blood glucose test will need to be repeated.