
Initial Startup
98
Installing or replacing the battery pack
When shipped, the battery pack is not installed in the
ACCU-CHEK Inform II meter.
After installing a new battery pack, the meter should be
charged for three hours in the base unit before testing.
Whenever the meter is in the base unit, the icon is
displayed. This icon shows that power is available and
the meter can charge, if necessary.
Make sure that the allowable temperature range for
charging the battery pack (37-104 °F or 3-40 °C) is
maintained during installation and initial setup.
Replace the battery pack within approximately 10
minutes to retain the date and time settings. Beyond
this period of time, you may have to re-enter the date
and time, or you may return the meter to the base unit
to obtain the correct date and time.
Kommentare zu diesen Handbüchern