Initial Startup
118
Unlocking a Download Lockout
When using a data management system for configura-
tion, a Download Lockout can be configured. This lockout
prevents a meter from being used for testing, if the meter
data have not been downloaded for a defined period of
time. If a download lockout occurs, the operator may use
a STAT test to temporarily override the lockout. If no STAT
test is configured or left, the meter cannot be used until
the stored data is downloaded via network (either wire-
less or by docking the meter).
In cases of e.g., network failures (which make a download
impossible), or other emergencies, it may be necessary to
unlock a meter without access to the data management
system. Since unlocking is a function of the Setup Menu,
it is restricted to persons who know the setup password.
To unlock a meter:
1 Touch in the Main Menu to open the Main
Menu 2 screen.
2 Touch Admin. to open the Admin Menu.
3 Touch Setup Menu to open the Setup Menu.
4 Enter the Setup password if requested.
5 Touch Unlock to reset the lockout timer.
The meter switches off, and is ready for testing after pow-
ering up again.
Main Menu 2 12:48pm
03/16/10
Maintenance
Strip Lots
Prociency
Control Lots
Beeper
Linearity
Diagnostics
Linearity Lots
Admin.
Admin Menu 12:48pm
03/16/10
Date / Time
Language Selection
Setup Menu
Setup Menu 12:48pm
03/16/10
Date / Time
QC
Test Display
Password
Test Ranges
Reagents
Operator ID Patient ID
Unlock