Patient Glucose Testing
42
Results screen The hourglass icon indicates the test is running. When
the test is completed and the result is ready, the meter
beeps again.
When the result is displayed, a message or warning may
also appear (depending on system configuration) notify-
ing you if the result exceeds the specified limit values.
These limit values define ranges that can either be con-
figured individually by the system administrator in line
with facility guidelines, or are the (technical) limits of the
system. The characteristics of these ranges are explained
on the following page.
The results screen contains a button that changes its
name based on the result (Range or Out of ... Range).
Touch this button to display the configured limit values.
Patient Test
03/16/10
12:48pm
Patient 123456789
Date 03/16/10 12:48
pm
150
I
Range
I-502:
Normal Range:
Critical Range:
Repor
table Range:
3.9-11.1 mmol/L
2.2-16.7 mmol/L
0.6-33.3 mmol/L
Patient Test 12:48pm
03/16/10
Patient 123456789
Strip Lot 545794
Patient Test
03/16/10
Range
8.3
12:48pm
Patient 123456789
Date 03/16/10 12:48
pm
mmol/L
Patient Test
03/16/10
Out of Reportable Range
RR HI
12:48pm
Patient 123456789
Date 03/16/10 12:48
pm
Patient Test
03/16/10
Out of Normal Range
12.2
12:48pm
Patient 123456789
Date 03/16/10 12:48
pm
mmol/L
Patient Test
03/16/10
Out of Critical Range
CR LO
12:48pm
Patient 123456789
Date 03/16/10 12:48
pm