Glucose Control Testing
56
Results screen The hourglass icon indicates the test is running. When
the test is completed and the result is ready, the meter
beeps again.
Depending on configuration, the result is displayed either
as a value or only as a qualitative result Pass or Fail. Your
system may be configured to disallow further testing until
all the required glucose control levels are successfully
run (QC Lockout).
When the results are displayed as a value, the results
screen contains a button that changes its name based on
the result (Range or Out of … Range). Touch this button
to display the minimum and maximum target values for
the control levels.
You can add comments to a test result (as with blood
glucose tests) (see page 45).
If you do not wish to add a comment to the test result,
touch to continue to the next level of the control
test, if necessary, or to return to the main menu.
Remove the test strip and dispose of it in accordance with
applicable regulations and directives.
Control Test
03/16/10
12:48pm
PASS
Control L2 (Hi) 123456
Date 03/16/10 12:48
pm
Control Test
03/16/10
Range
Control L2 (Hi) 123456
12:48pm
Date 03/16/10 12:48pm
17.1
mmol/L
Control Test
03/16/10
12:48pm
Control L2 (Hi) 123456
Date 03/16/10 12:48
pm
FAIL
Test results are also saved when the meter is powered
off in the results screen or when it automatically
powers off.