Proficiency Testing
90
Applying a proficiency sample Once the meter has detected the test strip, you are
prompted to apply the proficiency sample.
1 Wait until the flashing drop appears in the display
before applying the sample.
2 Apply a drop of sample to the front edge of the test
strip. Do not apply the sample to the top of the strip.
The sample is pulled into the test strip area by
capillary action.
Once sufficient sample has been detected, the meter
beeps and the measurement begins.
Prociency Test 12:48pm
03/16/10
Sample 123456789
Strip Lot 545794