Roche ACCU-CHEK Inform II Betreibershandbuch Seite 43

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 160
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 42
Patient Glucose Testing
43
The System Measurement Range refers to the
measurement range of the system itself (strips and
meter) and is the only range which cannot be
configured. With Accu-Chek Performa or
Accu-Chek Inform II test strips this fixed range is
10 to 600 mg/dL (0.6 to 33.3 mmol/L). If a result
falls outside this range, the message HI or LO
appears, i.e. , the result cannot be quantified
properly.
Glucose results above or below the Reportable
Range, as defined by the institution or local
regulatory body, are above the highest or below the
lowest numerical results to be reported. Results
outside this range must not be used for intervention
decisions.
Glucose results above or below the Critical Range,
as defined by the institution, require immediate
action as defined by hospital policy.
Glucose results within the Normal Range, as
defined by the institution, are considered normal
and require no therapeutic action.
Seitenansicht 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 159 160

Kommentare zu diesen Handbüchern

Keine Kommentare