Control Testing and Quality Control
110
Y
ou hear a beep when you have applied
enough control solution (provided the beeper
is enabled). The dropper icon disappears and
the test starts.
The result of the liquid quality control test is
displayed. It is automatically saved to memory.
The acceptable range of results for the liquid
control is displayed below the current result,
along with Pass or Fail.
If a quality control test fails, an up arrow (too
high) or down arrow (too low) is displayed and
flashes.
Note: The arrow (next to the result) refers to
the INR result only.
If you have selected to display INR and
%Quick or INR and seconds, the (up or down)
arrow next to the result refers only to the INR
value.
The printer icon only appears if the
printer function is activated. Otherwise it is not
displayed.
QC Test 09:15 am
04/19/2012
QC: 999
Code: 184
Op: Operator 1
Par: PT
Level: 1
QC Test 09:15 am
04/19/2012
1.2 INR
78 %Q
04/19/2012 09:15 am
QC: 999
Code: 184
Op: Operator 1
Par: PT
Level: 1
(1.0 - 1.4 INR)
(53 - 99 %Q)
Pass:
QC Test 09:15 am
04/19/2012
0.9 INRFail:
110 %Q
04/19/2012 09:15 am
QC: 999
Code: 184
Op: Operator 1
Par: PT
Level: 1
(1.0 - 1.4 INR)
(53 - 99 %Q)