Cleaning and Disinfecting the Meter
124
Cleaning/disinfecting the exterior (meter housing)
Use the solutions recommended on page 123 for cleaning/disinfecting the meter exterior. Apply
the solutions for a contact time of > 1 minute (refer to the corresponding product labeling).
Ensure that the blue test strip guide cover remains tightly closed while cleaning the meter
housing.
1 With the meter powered off, wipe the
meter's exterior clean.
Do not let liquid accumulate near any opening. Ensure that no liquid enters the meter.
2 With a fresh dry cloth or lint-free tissue,
wipe away residual moisture and fluids
after cleaning the housing.
3 Allow wiped areas to dry for at least
10 minutes before performing a test.