Operating conditions
T
o ensure that the meter functions properly, please observe the following guidelines:
■ Only use the meter at a room temperature between 15 °C and 32 °C (59 °F and 90 °F).
■ Only use the meter at a relative humidity between 10% and 85% (no condensation).
■ When operating the meter using the power adapter, use only a voltage of 100 V to 240 V
(± 10%) , 50/60 Hz.
■ When testing, place the meter on a level, vibration-free surface, or hold it so it is roughly
horizontal.
Quality control
The meter has a number of built-in quality-control functions:
■ A check of the electronic components and functions every time the meter is powered on.
■ A check of the test strip temperature while a test is in progress.
■ A check of the expiration date and lot information on the test strip based on the code chip
data.
■ A two-level, onboard quality control test and patient result determination within a single
test chamber.
Roche Diagnostics has available optional liquid quality controls for the CoaguChek XS Pro
system. These controls are provided to assist with meeting regulatory compliance requirements
as applicable to your facility.