Meter Setup
73
4 Select the triggers
, based on lot and/or
time, for requiring a quality control test.
You may select from the following options:
■ New Code Yes/No (applies every time a
new test strip lot is used)
The general intervals are:
■ No
■ Daily
■ Weekly
■ Monthly
5 Touch the button with the option of choice
when changing the test strip lot.
6 Touch the button to select the interval of
choice. Your selection is now highlighted.
7 Touch to save this setting, or
touch to exit this menu without
saving any changes.
8 For every option except for No, you must
now indicate the number of levels in
which the quality control must be per-
formed.
9 Touch to save this setting, or
touch to exit this menu without
saving any changes.
The display automatically returns to the
previous screen.
QC Lockout 09:15 am
04/19/2012
New Code
Yes No
Daily
W
eekly
Monthly
No
QC Lockout 09:15 am
04/19/2012
One Level
Two Levels
Number of Levels