Meter Setup
76
ST
AT Test Configuration
Quality control tests ensure consistent quality and accurate measurements. In emergency situa-
tions, however, it may be necessary to perform a test without delay. To override an active lock-
out
1
, you can allow for STAT Tests (STAT = Short TurnAround Time).
You can set the number of tests (up to a maximum of nine) permitted beyond the lockout. Once
the number of STAT Tests allowed has been reached, additional tests cannot be performed until
you have successfully completed a quality control test. The default setting is three STAT tests.
8 For every option except for No, you must
now indicate the number of levels in
which the quality control must be per-
formed.
9 Touch to save this setting, or
touch to exit this menu without
saving any changes.
The display automatically returns to the
previous screen.
1. QC, operator, or new lot lockout
The results of STAT tests are always identified by the label STAT Test when they are displayed
(see page 100).
Op. Lockout 09:15 am
04/19/2012
One Level
Two Levels
Number of Levels